Anatomical and Morphological Evaluation of Pacemaker Lead Compression
- 1 March 1993
- journal article
- editorial
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 16 (3), 434-444
- https://doi.org/10.1111/j.1540-8159.1993.tb01606.x
Abstract
In recent years, pacemaker lead failure due to compressive damage has been reported with increasing frequency. To document the mechanism of this failure, we evaluated explanted mechanically damaged leads with electrical testing, optical microscopy, and in some cases, scanning electron microscopy (SEM). In addition, we performed an autopsy study to measure the compressive loads on catheters placed percutaneously through the costoclavicular angle, as well as by cephalic cutdown. Of the 49 explanted compression damaged leads with enough clinical data for analysis, all had been placed by percutaneous subclavian puncture. Our autopsy data confirmed the significant increase in pressures generated in the costoclavicular angle for medial percutaneous subclavian catheterization (126 +/- 26 mmHg) compared to a more lateral percutaneous subclavian puncture (63 +/- 15 mmHg) or a cephalic cutdown (38 +/- 13 mmHg) (P < 0.01). In vivo coil compression testing documented loads up to 100 pounds per linear inch of coil and a compressive morphology by SEM identical to that seen in the clinical explants. Pacemaker leads appear to be susceptible to compression damage when placed by subclavian venipuncture. When possible, leads should be placed such that they avoid the tight costoclavicular angle.Keywords
This publication has 29 references indexed in Scilit:
- Survey of cardiac pacing in the United States in 1989The American Journal of Cardiology, 1992
- Facilitation of the Subclavian‐Introducer Technique with Contrast VenographyPacing and Clinical Electrophysiology, 1990
- Intermittent Oversensing due to Internal Insulation Damage of Temperature Sensing Rate Responsive Pacemaker Lead in Subclavian Venipuncture MethodPacing and Clinical Electrophysiology, 1989
- Anatomical bases for a safe method of subclavian venipunctureSurgical and Radiologic Anatomy, 1989
- Simultaneous Insulation Deterioration Associated with Side‐by‐Side Subclavian Placement of Two Polyurethane LeadsPacing and Clinical Electrophysiology, 1988
- Fluoroscopic Technique of Subclavian Venous Puncture for Permanent Pacing: A Safer and Easier ApproachPacing and Clinical Electrophysiology, 1988
- A Case of Pacemaker Lead Fracture Associated with Thoracic Outlet SyndromePacing and Clinical Electrophysiology, 1988
- Spontaneous intravenous catheter fracture and embolization from an implanted venous access port and analysis by scanning electron microscopyCancer, 1987
- Lead Fracture in Pacemaker PatientsThe Thoracic and Cardiovascular Surgeon, 1987
- Hickman Catheter SeparationJournal of Parenteral and Enteral Nutrition, 1985