A Prototype PET Scanner with DOI-Encoding Detectors
Open Access
- 13 June 2008
- journal article
- Published by Society of Nuclear Medicine in Journal of Nuclear Medicine
- Vol. 49 (7), 1132-1140
- https://doi.org/10.2967/jnumed.107.049791
Abstract
Detectors with depth-encoding allow a PET scanner to simultaneously achieve high sensitivity and high spatial resolution. Methods: A prototype PET scanner, consisting of depth-encoding detectors constructed by dual-ended readout of lutetium oxyorthosilicate (LSO) arrays with 2 position-sensitive avalanche photodiodes (PSAPDs), was developed. The scanner comprised 2 detector plates, each with 4 detector modules, and the LSO arrays consisted of 7 × 7 elements, with a crystal size of 0.9225 × 0.9225 × 20 mm and a pitch of 1.0 mm. The active area of the PSAPDs was 8 × 8 mm. The performance of individual detector modules was characterized. A line-source phantom and a hot-rod phantom were imaged on the prototype scanner in 2 different scanner configurations. The images were reconstructed using 20, 10, 5, 2, and 1 depth-of-interaction (DOI) bins to demonstrate the effects of DOI resolution on reconstructed image resolution and visual image quality. Results: The flood histograms measured from the sum of both PSAPD signals were only weakly depth-dependent, and excellent crystal identification was obtained at all depths. The flood histograms improved as the detector temperature decreased. DOI resolution and energy resolution improved significantly as the temperature decreased from 20°C to 10°C but improved only slightly with a subsequent temperature decrease to 0°C. A full width at half maximum (FWHM) DOI resolution of 2 mm and an FWHM energy resolution of 15% were obtained at a temperature of 10°C. Phantom studies showed that DOI measurements significantly improved the reconstructed image resolution. In the first scanner configuration (parallel detector planes), the image resolution at the center of the field of view was 0.9-mm FWHM with 20 DOI bins and 1.6-mm FWHM with 1 DOI bin. In the second scanner configuration (detector planes at a 40° angle), the image resolution at the center of the field of view was 1.0-mm FWHM with 20 DOI bins and was not measurable when using only 1 bin. Conclusion: PET scanners based on this detector design offer the prospect of high and uniform spatial resolution (crystal size, ∼1 mm; DOI resolution, ∼2 mm), high sensitivity (20-mm-thick detectors), and compact size (DOI encoding permits detectors to be tightly packed around the subject and minimizes number of detectors needed).Keywords
This publication has 34 references indexed in Scilit:
- Measurements of wavelength shifting (WLS) fibre readout for a highly multiplexed, depth-encoding PET detectorPhysics in Medicine & Biology, 2007
- Depth of interaction decoding of a continuous crystal detector modulePhysics in Medicine & Biology, 2007
- Monte Carlo simulations of absorbed dose in a mouse phantom from 18‐fluorine compoundsMedical Physics, 2007
- Performance evaluation of the ECAT HRRT: an LSO-LYSO double layer high resolution, high sensitivity scannerPhysics in Medicine & Biology, 2007
- Position Sensitive APDs for Small Animal PET ImagingIEEE Transactions on Nuclear Science, 2004
- A prototype modular detector design for high resolution positron emission mammography imagingIEEE Transactions on Nuclear Science, 2003
- Dual APD array readout of LSO crystals: optimization of crystal surface treatmentIEEE Transactions on Nuclear Science, 2002
- Performance evaluation of the microPET P4: a PET system dedicated to animal imagingPhysics in Medicine & Biology, 2001
- Investigation of depth-of-interaction by pulse shape discrimination in multicrystal detectors read out by avalanche photodiodesIEEE Transactions on Nuclear Science, 1999
- A room temperature LSO/PIN photodiode PET detector module that measures depth of interactionIEEE Transactions on Nuclear Science, 1995