Lens Injury Following Iridotomy With a Q-Switched Neodymium-YAG Laser
- 1 January 1986
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 104 (1), 123-125
- https://doi.org/10.1001/archopht.1986.01050130137038
Abstract
• By scanning electron microscopy, we studied eight lenses from eyes that had received a neodymium-YAG laser iridotomy one day prior to elective intracapsular cataract extraction. In some of these cases, pitting lesions were found on the anterior-equatorial lens surface, thereby showing that lenticular damage may occur following this procedure if the iridotomy is applied too centrally. It is recommended that care be taken to ensure that the iridotomy be placed at a site peripheral to the outer circumference of the lens.This publication has 12 references indexed in Scilit:
- Argon Laser-Induced Cataract as a Complication of Retinal PhotocoagulationArchives of Ophthalmology (1950), 1982
- Photocoagulation-Induced Lens OpacityArchives of Ophthalmology (1950), 1982
- Argon Laser Peripheral Iridotomies in the Treatment of Primary Angle Closure GlaucomaArchives of Ophthalmology (1950), 1982
- Clinical studies on the efficiency of high power laser radiation upon some structures of the anterior segment of the eyeInternational Ophthalmology, 1981
- Protocol for Single-Session Argon Laser Iridectomy for Angle-Closure GlaucomaInternational Ophthalmology Clinics, 1981
- Argon Laser IridectomiesAmerican Journal of Ophthalmology, 1973
- Histopathology and Ultrastructure of the Argon Laser Lesion in Human Retinal and Choroidal VasculaturesAmerican Journal of Ophthalmology, 1973
- Improvements in Technique of Photocoagulation of the IrisArchives of Ophthalmology (1950), 1965
- Laser Coagulation of Ocular TissuesArchives of Ophthalmology (1950), 1964
- Photocoagulation of Iris and RetinaArchives of Ophthalmology (1950), 1958