Dependence of Transcutaneous Oxygen Tension on Local Arteriovenous Pressure Gradient in Normal Subjects

Abstract
1. We studied the relationship between trans-cutaneous oxygen tension at the foot and local arteriovenous pressure difference in 15 normal men and women; arteriovenous pressure difference was varied by changing the height of the foot with respect to the heart and by applying external pressure to the foot. 2. Control transcutaneous oxygen tension was 67 ± 9 sd mmHg (8.9 ± 1.2 kPa) at a control arteriovenous pressure difference of 80 ± 6 sd mmHg (10.6 ± 0.8 kPa). 3. In every subject transcutaneous oxygen tension fell non-linearly with a decrease in arteriovenous pressure difference; transcutaneous oxygen tension was relatively insensitive to changes in arteriovenous pressure difference when arteriovenous pressure difference was high, but always fell sharply to zero at some positive arteriovenous pressure difference [range 13-34 mmHg (1.7-4.5 kPa)]. 4. An analysis of the data indicated that transcutaneous oxygen tension varied with arteriovenous pressure difference approximately as the oxygen tension of cutaneous venous blood under the sensor varied (in the absence of changes in cutaneous vascular resistance and oxygen consumption). 5. This analysis was supported by studies in three subjects in whom the oxygen tension of superficial venous drainage from a warmed hand or foot was measured along with Transcutaneous oxygen tension while arteriovenous pressure difference was varied.