Context.—A rapid, real-time, duplex, fluorescent molecular beacon probe–based polymerase chain reaction (PCR) assay was recently developed for the detection of methicillin-resistant Staphylococcus aureus. Objective.—To describe the development and validation of this unique assay. Design.—Prospective laboratory analysis. Setting.—Urban health region/centralized diagnostic microbiology laboratory. Bacterial Strains.—One hundred eighty-one previously characterized clinical and American Type Culture Collection isolates, including 50 strains each of methicillin-resistant and methicillin-sensitive S aureus, plus 50 strains of coagulase-negative staphylococci and 31 nonstaphylococcal isolates to ensure assay specificity. Intervention.—Assays were performed on purified genomic DNA extracted from growing bacterial colonies. Two sets of oligonucleotide primers were used to specifically amplify the mecA and nuc genes, followed by detection of amplicons using fluorophore-labeled molecular beacon probes. Assays were performed on the Mx4000 Multiplex Quantitative PCR System (Stratagene Inc, La Jolla, Calif). Main Outcome Measures.—(1) Assay sensitivity and specificity, and (2) analytical sensitivity. Results.—The assay demonstrated 100% sensitivity and 100% specificity, and accurately characterized isolates as methicillin-resistant S aureus, methicillin-sensitive S aureus, or methicillin-resistant coagulase-negative staphylococci, with test results available in 2.5 hours. The analytical sensitivity of the assay was determined to be between 6 and 60 genomic equivalents. Conclusions.—This assay is rapid, accurate, easy to perform, and is compatible with other real-time PCR instruments, making it a suitable alternative to conventional PCR methodologies.Keywords
This publication has 15 references indexed in Scilit:
- Factors influencing methicillin resistance in staphylococciArchiv für Mikrobiologie, 2002
- Detection of Methicillin-Resistant Staphylococcus aureus and Simultaneous Confirmation by Automated Nucleic Acid Extraction and Real-Time PCRJournal of Clinical Microbiology, 2002
- Rapid Identification of Staphylococcus aureus and the mecA Gene from BacT/ALERT Blood Culture Bottles by Using the LightCycler SystemJournal of Clinical Microbiology, 2002
- Real-Time Multiplex PCR AssaysMethods, 2001
- Rapid Identification of Methicillin-Resistant Staphylococcus aureus from Positive Blood Cultures by Real-Time Fluorescence PCRJournal of Clinical Microbiology, 2001
- Containing methicillin-resistantS aureusPostgraduate Medicine, 2001
- MRSA patients: proven methods to treat colonization and infectionJournal of Hospital Infection, 2001
- The Progressive Intercontinental Spread of Methicillin-Resistant Staphylococcus aureusClinical Infectious Diseases, 1997
- Real time quantitative PCR.Genome Research, 1996
- Molecular Beacons: Probes that Fluoresce upon HybridizationNature Biotechnology, 1996