Measurement of Resolving Times with the Fast RCA C-70045 Photomultipliers
- 1 June 1966
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 13 (3), 127-131
- https://doi.org/10.1109/TNS.1966.4324088
Abstract
This work deals with tests of the RCA C-70045 photomultipliers in a delayed coincidence apparatus. Illuminating the photocathode of the photomultiplier with constant height light pulses, resolving time measurements are performed. Prompt resolution curves are then measured using two photomultipliers of this type associated with two plastic scintillators. A time resolution of 1.85×1O-10s (f.w. h.m.) was obtained with Co60.Keywords
This publication has 6 references indexed in Scilit:
- Zeitauflösungseigenschaften von NaJ-SzintillatorenNuclear Instruments and Methods, 1965
- Fast delayed coincidence technique: The XP1020 photomultiplier and limits of resolving times due to scintillator characteristicsNuclear Instruments and Methods, 1964
- Revised theory of time resolution in scintillation countersNuclear Instruments and Methods, 1964
- Evaluation of the C-70045A High-Speed PhotomultiplierIEEE Transactions on Nuclear Science, 1964
- A survey of the latest developments in delayed coincidence measurementsNuclear Instruments and Methods, 1963
- Theory of time resolution in scintillation countersNuclear Instruments and Methods, 1959