Validation of Doppler-derived pulmonary arterial pressure in patients with ductus arteriosus under different hemodynamic states.
- 1 November 1987
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 76 (5), 1081-1091
- https://doi.org/10.1161/01.cir.76.5.1081
Abstract
Twenty-nine patients with a patent ductus arteriosus (PDA) in isolation (n = 17) or in combination with other lesions (n = 12) underwent simultaneous hemodynamic assessment and evaluation of PDA flow velocity by the Doppler method. The accuracy with which Doppler velocity across the PDA predicted pulmonary arterial pressure and the influence of PDA size and shape on the Doppler velocity-pressure relationship were examined. Seventy percent had a cone-shaped PDA (narrowest at the pulmonary artery end), and the remainder were tubular. Narrowest PDA diameter ranged from 1.5 to 9 mm (mean 3.5 mm). Peak systolic and mean pulmonary arterial pressure ranged from 10 to 116 and 8 to 72 mm Hg, respectively. Twenty-one patients (group 1) had left-to-right shunting only. The following variables showed significant correlation in this group: peak instantaneous systolic aortic-to-main pulmonary arterial (MPA) pressure gradient and maximum Doppler velocity across the PDA (slope = 1.03, SEE = 13 mm Hg, r = .94, p less than .001), mean aortic-to-MPA pressure gradient and mean Doppler velocity (slope = 1.06, SEE = 10 mm Hg, r = .95, p less than .001), and end diastolic aortic-to-MPA pressure gradient and minimum Doppler velocity (slope = 1.12, SEE = 8 mm Hg, r = .96, p less than .001). Eight patients (group 2) had bidirectional shunting. In this group peak instantaneous aortic-to-MPA pressure gradient significantly correlated with maximum Doppler velocity measured from the left-to-right shunt (slope = .70, SEE = 2 mm Hg, r = .92, p less than .002) and mean pressure gradient correlated with mean Doppler velocity (slope = .83, SEE = 3 mm Hg, r = .78, p less than .003). Right-to-left Doppler velocities showed no correlation with pressures. In six patients with pulmonary hypertension Doppler velocity changes accurately predicted the effect of pulmonary vasodilation on pulmonary arterial pressure. Doppler velocity of PDA flow reliably predicts pulmonary arterial pressure over a wide range of pressures and PDA shapes and sizes.This publication has 7 references indexed in Scilit:
- Doppler echocardiographic estimation of systolic pulmonary artery pressure in patients with aortic-pulmonary shuntsJournal of the American College of Cardiology, 1986
- The accuracy of Doppler ultrasound measurement of pressure gradients across irregular, dual, and tunnellike obstructions to blood flow.Circulation, 1985
- Assessment of the ductus arteriosus in preterm infants utilizing suprasternal two-dimensional/doppler echocardiographyJournal of the American College of Cardiology, 1985
- Combined noninvasive assessment of the patent ductus arteriosus in the preterm infant before and after indomethacin treatmentThe American Journal of Cardiology, 1984
- Factors affecting accuracy of in vitro valvar pressure gradient estimates by doppler ultrasoundThe American Journal of Cardiology, 1984
- In vitro verification of doppler prediction of transvalve pressure gradient and orifice area in stenosisThe American Journal of Cardiology, 1984
- Suprasternal cross-sectional echocardiography in assessment of patient ducts arteriosus.Heart, 1982