Resonance frequency analysis in relation to jawbone characteristics and during early healing of implant installation

Abstract
Objectives: To monitor resonance frequency analysis (RFA) in relation to the jawbone characteristics and during the early phases of healing and incorporation of Straumann® dental implants with an SLA surface. Material and methods: 17 Straumann 4.1 mm implants (10 mm) and 7 Straumann 4.8 mm implants (10 mm) were installed and ISQ determined at baseline and after 1, 2, 3, 4, 5, 6, 8 and 12 weeks. Central bone cores were analyzed from the 4.1 mm implants using micro CT for bone volume density (BVD) and bone trabecular connectivity (BTC). Results: Pocket probing depths ranged from 2–4 mm and bleeding on probing from 5–20%. At baseline, BVD varied between 24% and 65% and BTC between 4.9 and 25.4 for the 4.1 mm implants. Baseline ISQ varied between 55 and 74 with a mean of 61.4. No significant correlations were found between BVD or BTC and ISQ Values. For the 4.8 mm diameter implants baseline ISQ values ranged from 57–70 with a mean of 63.3. Over the healing period ISQ values increased at 1 week and decreased after 2–3 weeks. After 4 weeks ISQ values, again increased slightly, no significant differences were noted over time. One implant (4.1 mm) lost stability at 3 weeks. Its ISQ value had dropped from 68 to 45. However the latter value was determined after the clinical diagnosis of instability. Conclusion: ISQ values of 57–70 represented homeostasis and implant stability. However no predictive value for loosing implant stability can be attributed to RFA since the decrease occurred after the fact.