Minimizing Electrode Motion Artifact by Skin Abrasion
- 1 March 1977
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-24 (2), 134-139
- https://doi.org/10.1109/tbme.1977.326117
Abstract
Motion artifacts frequently cause erroneous results in biopotential measurement. When surface electrodes and conducting paste are used, a major cause of motion artifact is change in skin potential. The most effective means of minimizing this artifact is light abrasion of the sidn with fine sandpaper. A very mild paste is then required to avoid skin irritation.Keywords