Monitoring of Intracellular Ca2+ Elevation in a Single Neural Cell Using a Fluorescence Microscope/Video-Camera System

Abstract
For monitoring the changes in intracellular free Ca2+ concentration ([Ca2+]i), we developed a simple system combining a fluorescence microscope, an image intensifier, a video-camera, a cathode ray tube display and a photodiode, employing quin2 as a Ca2+ indicator. We recorded increases of the fluorescence intensity due to [Ca2+]i rises, when high K+ medium, neurotransmitter and Ca2+ ionophore were applied to the single cells of nervous system origin in culture. The present system is capable of simultaneous detection of the [Ca2+]i changes from multiple separate cells.