A reliable method for localizing deep intracranial sources of the EEG
- 1 December 1985
- journal article
- research article
- Published by Wolters Kluwer Health in Neurology
- Vol. 35 (12), 1702
- https://doi.org/10.1212/wnl.35.12.1702
Abstract
We have demonstrated the reliability of a noninvasive method for successfully localizing the intracranial origin of the EEG. The dipole localization method (DLM) is a computer-assisted, mathematical method based on electrical field theory and is similar to localization methods currently used by electrocardiologists. In 12 patients with intractable epilepsy who were being evaluated for surgery, a known current source was introduced between two adjacent depth electrodes. Using scalprecorded EEG only, DLM accurately and reliably localized the source to within 2 cm of the known origin in all instances where a discrete source was present. We conclude that DLM is a valid and reliable noninvasive method for localizing the intracranial source of some scalp-recorded EEG potentials, and that in some patients, use of this method may obviate the need for depth electrode implantation.This publication has 3 references indexed in Scilit:
- Scalp and depth recordings of induced deep cerebral potentialsElectroencephalography and Clinical Neurophysiology, 1983
- Electrophysiological studies of long-term electrical stimulation of the cerebellum in monkeysJournal of Neurosurgery, 1977
- The electrical fields at the surface of the head during sleepElectroencephalography and Clinical Neurophysiology, 1949