Evaluation of the sensititre microdilution antibiotic susceptibility system against recent clinical isolates: three-laboratory collaborative study
- 1 April 1980
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 11 (4), 426-429
- https://doi.org/10.1128/jcm.11.4.426-429.1980
Abstract
The Sensititre broth microdilution antibiotic susceptibility testing system [for bacteria] was comparable to a reference broth dilution method. Of 3194 minimum inhibitory concentration data pairs, 93.9% were within the acceptable .+-.1-log2 dilution range. Only rare (0.6%) very major errors were found between methods.This publication has 4 references indexed in Scilit: