Quantitative evaluation of several partial fourier reconstruction algorithms used in mri
- 1 July 1993
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 30 (1), 51-59
- https://doi.org/10.1002/mrm.1910300109
Abstract
Partial Fourier reconstruction algorithms exploit the redundancy in magnetic resonance data sets so that half of the data is calculated during image reconstruction rather than acquired. The conjugate synthesis, Margosian, homodyne detection, Cuppen and POCS algorithms are evaluated using spatial frequency domain analysis to show their characteristics and where limitations may occur. The phase correction used in partial Fourier reconstruction is equivalent to a convolution in the frequency domain and the importance of accurately implementing this convolution is demonstrated. New reconstruction approaches, based on passing the partial data through a phase correcting, finite impulse response (FIR), digital filter are suggested. These FIR and MoFIR algorithms have a speed near that of the Margosian and homodyne detection reconstructions, but with a lower error; close to that of the Cuppen/POCS iterative approaches. Quantitative analysis of the partial Fourier algorithms, tested with three phase estimation techniques, are provided by comparing artificial and clinical data reconstructed using full and partial Fourier techniques.Keywords
This publication has 8 references indexed in Scilit:
- Noise and artifact comparison for Fourier and polynomial phase correction used with Fourier reconstruction of asymmetric data setsJournal of Magnetic Resonance Imaging, 1992
- Homodyne detection in magnetic resonance imagingIEEE Transactions on Medical Imaging, 1991
- A quantitative comparison of the TERA modeling and DFT magnetic resonance image reconstruction techniquesMagnetic Resonance in Medicine, 1991
- A fast, iterative, partial-fourier technique capable of local phase recoveryJournal of Magnetic Resonance (1969), 1991
- Improved detectability in low signal‐to‐noise ratio magnetic resonance images by means of a phase‐corrected real reconstructionMedical Physics, 1989
- Correction of spatially dependent phase shifts for partial Fourier imagingMagnetic Resonance Imaging, 1988
- Reducing MR imaging time by one-sided reconstructionMagnetic Resonance Imaging, 1987
- Halving MR imaging time by conjugation: demonstration at 3.5 kG.Radiology, 1986