Fast 2D phantom dosimetry for scanning proton beams
- 1 April 1998
- journal article
- Published by Wiley in Medical Physics
- Vol. 25 (4), 464-475
- https://doi.org/10.1118/1.598221
Abstract
A quality control system especially designed for dosimetry in scanning proton beams has been designed and tested. The system consists of a scintillating screen (Gd2O2S:Tb), mounted at the beam-exit side of a phantom, and observed by a low noise CCD camera with a long integration time. The purpose of the instrument is to make a fast and accurate two-dimensional image of the dose distribution at the screen position in the phantom. The linearity of the signal with the dose, the noise in the signal, the influence of the ionization density on the signal, and the influence of the field size on the signal have been investigated. The spatial resolution is 1.3 mm (1 s.d.), which is sufficiently smaller than typical penumbras in dose distributions, The measured yield depends linearly on the dose and agrees within 5% with the calculations. In the images a signal to noise ration (signal/l s.d.) of 10(2) has been found, which is in the same order of magnitude as expected from the calculations. At locations in the dose distribution possessing a strong contribution of high ionization densities (i.e., in the Bragg peak), we found some quenching of the light output, which can be described well by existing models if the beam characteristics are known. For clinically used beam characteristics such as a Spread Out Bragg peak, there is at most 8% deviation from the NACP ionization chamber measurements. The conclusion is that this instrument is a useful tool for quick and reliable quality control of proton beams. The long integration-time capabilities of the system make it worthwhile to investigate its applicability in scanning proton beams and other dynamic treatment modalities. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)02104-X]Keywords
This publication has 30 references indexed in Scilit:
- Two-dimensional exit dosimetry using a liquid-filled electronic portal imaging device and a convolution modelRadiotherapy and Oncology, 1997
- Transmission dosimetry with a liquid-filled electronic portal imaging deviceInternational Journal of Radiation Oncology*Biology*Physics, 1996
- The use of an electronic portal imaging device for exit dosimetry and quality control measurementsInternational Journal of Radiation Oncology*Biology*Physics, 1995
- The 200‐MeV proton therapy project at the Paul Scherrer Institute: Conceptual design and practical realizationMedical Physics, 1995
- Measurement possibilities using an electronic portal imaging deviceRadiotherapy and Oncology, 1993
- NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: A new approach to 3D dosimetry by MRIMagnetic Resonance Imaging, 1993
- A review of electronic portal imaging devices (EPIDs)Medical Physics, 1992
- MR imaging of absorbed dose distributions for radiotherapy using ferrous sulphate gelsPhysics in Medicine & Biology, 1990
- A digital fluoroscopic imaging device for radiotherapy localizationInternational Journal of Radiation Oncology*Biology*Physics, 1990
- Film dosimetry of clinical electron beamsInternational Journal of Radiation Oncology*Biology*Physics, 1990