Nondispersive Infrared Spectrometry: A New Method for the Detection ofHelicobacter pyloriInfection with the 13C‐Urea Breath Test
Open Access
- 1 November 1997
- journal article
- Published by Oxford University Press (OUP) in Clinical Infectious Diseases
- Vol. 25 (5), 1003-1005
- https://doi.org/10.1086/516069
Abstract
Nondispersive infrared spectrometry (NDIRS) was used to detect Helicobacter pylori infection with the 13C-urea breath test. The results were compared with those of standard isotope ratio mass spectrometry (IRMS). Both methods accurately distinguished between H. pylori-positive and H. pylori-negative individuals. The results demonstrate that NDIRS technology is accurate and therefore of equal value to standard IRMS for detection of H. pylori infection. It can be recommended for routine clinical application. As NDIRS technology is much cheaper than current IRMS machines, we consider the new method extremely useful for clinical applications.Keywords
This publication has 2 references indexed in Scilit:
- Helicobacter pylori: Current statusGastroenterology, 1993
- Effect of Ranitidine and Amoxicillin plus Metronidazole on the Eradication of Helicobacter pylori and the Recurrence of Duodenal UlcerNew England Journal of Medicine, 1993