Pressure Monitoring Devices
- 23 August 1976
- journal article
- research article
- Published by American Medical Association (AMA) in JAMA
- Vol. 236 (8), 936-938
- https://doi.org/10.1001/jama.1976.03270090030022
Abstract
Contaminated pressure monitoring devices recently have been implicated as a source of epidemic organisms in three outbreaks of nosocomial bacteremia, one outbreak of candidemia, and one outbreak of hepatitis. Measures necessary to prevent monitoring-related infections have not always been appreciated or taken. As a minimum, pressure monitoring devices should be sterilized between use with different patients; strict aseptic technique should be employed when setting up and using monitoring systems; and each patient's monitoring tubing, fluid, and monitoring devices should be changed at regular intervals. (JAMA236:936-938, 1976)Keywords
This publication has 3 references indexed in Scilit:
- Percutaneous Indwelling Radial-Artery Catheters for Monitoring Cardiovascular FunctionNew England Journal of Medicine, 1974
- Sterility of fluid–filled lines for pressure measurementAnaesthesia, 1972
- CONTROL OF HEPATITIS IN DIALYSIS UNITSBritish Medical Bulletin, 1972