Measurement of breath amplitudes: Comparison of three noninvasive respiratory monitors to integrated pneumotachograph
- 21 October 1993
- journal article
- research article
- Published by Wiley in Pediatric Pulmonology
- Vol. 16 (4), 254-258
- https://doi.org/10.1002/ppul.1950160408
Abstract
Measurement of breath amplitude (BA) and similarly tidal volume (VT) in newborn infants is the standard for detection of apnea and hypopneas. The purpose of our study was to compare the accuracy for BA by three frequently utilized noninvasive respiratory monitors: respiratory inductive plethysmography (RIP), mercury in silastic strain gauges (SG), and impedance pneumography (IP). Twenty healthy full‐term infants were studied in both supine and prone postures. The RC and AB gain factors for RIP were obtained using qualitative diagnostic calibration (QDC) procedure. The electrical gain of IP was set equivalent to the BA signal of a pneumotachograph (PNT). The three devices were calibrated in the supine posture and measurements were repeated in the prone posture without changing their calibration factors. Compared to PNT, postural change did not significantly alter BA measured by RIP. The accuracy of breath‐to‐breath BA measurement in the prone posture was worse for IP and SG compared to RIP and PNT. In contrast to SG or IP, the accuracy of BA measurement maintained was by RIP after a postural change from supine to prone in fullterm newborns. Pediatr Pulmonol. 1993; 16:254–258.Keywords
This publication has 9 references indexed in Scilit:
- Tidal Volume Measurements in Newborns Using Respiratory Inductive PlethysmographyAmerican Review of Respiratory Disease, 1993
- Calibration of respiratory inductive plethysmograph during natural breathingJournal of Applied Physiology, 1989
- Simultaneous Quantification of Chest Wall Distortion by Multiple Methods in Preterm InfantsAmerican Review of Respiratory Disease, 1988
- Comparison of respiratory inductive plethysmography and thoracic impedance for apnea monitoringThe Journal of Pediatrics, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Impedance pneumograph and magnetometer methods for monitoring tidal volume.Journal of Applied Physiology, 1974
- Measurement of the separate volume changes of rib cage and abdomen during breathingJournal of Applied Physiology, 1967
- COMPARISON OF METHODS FOR RECORDING RESPIRATION IN NEWBORN INFANTSPediatrics, 1965