Abstract
A novel, solid-supported voltammetric ion-selective electrode to detect anticoagulant/antithrombotic heparin at polarizable poly(vinyl chloride) (PVC) membrane/water interfaces was developed. An ∼3−4.5-μm-thick PVC membrane plasticized with 2-nitrophenyl octyl ether was supported on a gold electrode modified with a poly(3-octylthiophene) (POT) film as an ion-to-electron transducer. Charge transport through the PVC-covered POT film is electrochemically reversible, as demonstrated by cyclic voltammetry with nonpolarizable membrane/water interfaces. In addition to the fast charge transport, adequate redox capacity of the POT film and a small ohmic potential drop in the thin PVC membrane enable ion transfer voltammetry at polarizable macroscopic membrane/water interfaces in a standard three-electrode cell. Reversible ClO4- transfer at the interfaces coupled with oxidation of a neutral POT film was examined by cyclic voltammetry to determine the distribution of the applied potential to the two polarizable interfaces by convolution technique. Interfacial adsorption and desorption of heparin facilitated by octadecyltrimethylammonium were studied also by cyclic voltammetry and convolution technique to demonstrate that the processes are electrochemically irreversible. Stripping voltammetry based on the interfacial processes gives a low detection limit of 0.005 unit/mL heparin in a saline solution, which is slightly lower than the detection limit of most sensitive heparin sensors reported so far (0.01 unit/mL).