Planar Amperometric Glucose Sensor Based on Glucose Oxidase Immobilized by Chitosan Film on Prussian Blue Layer
Open Access
- 22 April 2002
- Vol. 2 (4), 127-136
- https://doi.org/10.3390/s20400127
Abstract
A planar amperometric glucose microsensor based on glucose oxidase immobilized by chitosan film on Prussian Blue layer has been developed. The experimental results show that the optimum detection potential is 50 mV (versus Ag/AgCl) and the optimum pH is 6.5. Under the selective conditions the sensor exhibits excellent sensitivity of 98 nA/M and a linear range of 0.1-6.0 mM. The apparent Michanelis-Menten constant of the sensor is 21 mM. The response time is less than 60 seconds. No apparent change in the response to glucose was observed during one month. Foremost, the interference of ascorbic and uric acids can be avoided due to selective permeability of chitosan film and electrocatalysis of PB layer to H2O2. The sensor has been applied to detect glucose in human blood serum.Keywords
This publication has 18 references indexed in Scilit:
- Evaluation of glucose biosensors based on Prussian Blue and lyophilised, crystalline and cross-linked glucose oxidases (CLEC®)Talanta, 2001
- Improvement in the performance of a nickel complex-based electrochemical sensor for the detection of nitric oxide in solutionSensors and Actuators B: Chemical, 1999
- On the mechanism of H2O2 reduction at Prussian Blue modified electrodesElectrochemistry Communications, 1999
- Glucose oxidase/polyion complex-bilayer membrane for elimination of electroactive interferents in amperometric glucose sensorAnalytica Chimica Acta, 1998
- Strategies for decreasing ascorbate interference at glucose oxidase-modified poly(o-phenylenediamine)-coated electrodesThe Analyst, 1996
- Prussian Blue-Based First-Generation Biosensor. A Sensitive Amperometric Electrode for GlucoseAnalytical Chemistry, 1995
- Effect of interference on amperometric glucose biosensors with cellulose acetate membranesElectroanalysis, 1994
- Electropolymerized 1,2-diaminobenzene as a means to prevent interferences and fouling and to stabilize immobilized enzyme in electrochemical biosensorsAnalytical Chemistry, 1990
- Catalysis of the reduction of molecular oxygen to water at Prussian blue modified electrodesJournal of the American Chemical Society, 1984