Tonometry After Laser in Situ Keratomileusis Treatment

Abstract
To assess the reliability of intraocular pressure measurements by Goldmann applanation tonometry versus pneumotonometry after laser in situ keratomileusis for myopia. In this prospective study, corneal Goldmann applanation tonometry and pneumotonometry measurements were made in 118 eyes of 60 patients before and 1 and 3 months after undergoing laser in situ keratomileusis for myopia. Manifest refraction, ultrasonic corneal thickness measurements, and keratometry readings were also obtained. Preoperative intraocular pressure showed a good correlation between Goldmann applanation tonometry and pneumotonometry values (Pearson r = 0.71;P P Contact pneumotonometry measures the IOP reliably after laser in situ keratomileusis for myopia, whereas Goldmann applanation tonometry underestimates the intraocular pressure. This may be important in the treatment of any future glaucoma.