Multisection-multiecho MR imaging: effect on image quality.

Abstract
An analysis of the effects of multisection-multiecho imaging on signal intensity is presented using a gaussian excitation pulse. The number of sections acquired, size of the intersection gap, and the number of echoes were all varied independently to evaluate their effect on image intensity. The results indicate that one should take a conservative approach in multisection-multiecho imaging, leaving at least a 2-mm-intersection gap and approximately a 20-msec delay between section excitations to avoid image signal loss owing to partial saturation effects.