Variations in radiation dose between the same model of multislice CT scanner at different hospitals
- 1 November 2003
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 76 (911), 798-802
- https://doi.org/10.1259/bjr/33117342
Abstract
The variation in exposure factors and patient dose, between seven centres using identical multislice CT scanners, was investigated for six standard examinations. Dose values were compared with each other and the relevant diagnostic reference level (DRL) for each examination. The range in weighted CT dose index (CTDI(w)) values between the seven centres was small for abdominal scans and head scans. For other scans however, such as functional endoscopic sinonasal surgery (FESS) the variations in CTDI(w) were as high as a factor of seven between the lowest and the highest values. At one centre a program of dose optimization had been undertaken and this centre had CTDI(w) values ranging from 3% to 64% lower than the average value for the seven centres. This demonstrates that significant dose reduction can be achieved through close collaboration between medical physicists, radiologists and radiographers.Keywords
This publication has 7 references indexed in Scilit:
- A method to obtain the same levels of CT image noise for patients of various sizes, to minimize radiation doseThe British Journal of Radiology, 2002
- Radiation dose in CT: are we meeting the challenge?The British Journal of Radiology, 2002
- A survey of patient dose and image quality for computed tomography scanners in WalesJournal of Radiological Protection, 2001
- CT scanning: patterns of use and doseJournal of Radiological Protection, 2000
- Application of draft European Commission reference levels to a regional CT dose survey.The British Journal of Radiology, 2000
- Spiral CT: how much does radiation dose matter?The Lancet, 1998
- CT scanner dosimetry.The British Journal of Radiology, 1998