Abstract
High-frequency alternating current (AC) waveforms have been shown to produce a quickly reversible nerve block in animal models, but the parameters and mechanism of this block are not well understood. A frog sciatic nerve/gastrocnemius muscle preparation was used to examine the parameters for nerve conduction block in vivo, and a computer simulation of the nerve membrane was used to identify the mechanism for block. The results indicated that a 100% block of motor activity can be accomplished with a variety of waveform parameters, including sinusoidal and rectangular waveforms at frequencies from 2 kHz to 20 kHz. A complete and reversible block was achieved in 34 out of 34 nerve preparations tested. The most efficient waveform for conduction block was a 3–5 kHz constant-current biphasic sinusoid, where block could be achieved with stimulus levels as low as 0.01 μC phase−1. It was demonstrated that the block was not produced indirectly through fatigue. Computer simulation of high-frequency AC demonstrated a steady-state depolarisation of the nerve membrane, and it is hypothesised that the conduction block was due to this tonic depolarisation. The precise relationship between the steady-state depolarisation and the conduction block requires further analysis. The results of this study demonstrated that high-frequency AC can be used to produce a fast-acting, and quickly reversible nerve conduction block that may have multiple applications in the treatment of unwanted neural activity.Keywords
This publication has 55 references indexed in Scilit:
- REDUCTION OF BLADDER OUTLET RESISTANCE BY SELECTIVE SACRAL ROOT STIMULATION USING HIGH-FREQUENCY BLOCKADE IN DOGSJournal of Urology, 1998
- Inversion of the current-distance relationship by transient depolarizationIEEE Transactions on Biomedical Engineering, 1997
- Stimulator design and subsequent stimulation parameter optimization for controlling micturition and reducing urethral resistanceIEEE Transactions on Rehabilitation Engineering, 1996
- Long-Term Effect of Sphincteric Fatigue During Bladder NeurostimulationJournal of Urology, 1995
- Cyclic AMP response in cells exposed to electric fields of different frequencies and intensitiesRadiation and Environmental Biophysics, 1994
- Dose-response study of the pathological effects of chronically applied direct current stimulation on the normal rat spinal cordJournal of Neurosurgery, 1993
- Local anaesthetic block protects against electrically-induced damage in peripheral nerveJournal of Biomedical Engineering, 1990
- Response of Single Alpha Motoneurons to High-Frequency Pulse TrainsStereotactic and Functional Neurosurgery, 1986
- Histopathological Evaluation of Dog Sacral Nerve after Chronic Electrical Stimulation for MicturitionNeurosurgery, 1984
- Reversible Blocking of Nerve Conduction by Alternating-Current ExcitationNature, 1962