Effects of structural configuration of prosthetic aortic valves upon coronary blood flow
- 1 August 1977
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Surgery
- Vol. 64 (8), 561-566
- https://doi.org/10.1002/bjs.1800640810
Abstract
This study was performed in order to determine the effects of the structural configuration of prosthetic aortic valves upon coronary blood flow. A pulse-duplicating system was utilized to which a rigid model of the root of the aorta and the left artery were attached. A stented normal human valve, stented porcine valve, pivoting disc valve, tilting disc valve and a caged ball valve were tested at simulated resting conditions. The results showed that the pulse configuration and proportions of coronary flow that were systolic and diastolic changed with various valves. There was no significant reduction of the mean simulated coronary flow with the porcine valve or disc valves. Coronary flow was augmented with the caged ball valve. These differences were interpreted as being due to several fluid dynamic phenomena. These were piston effects, inertia of reversed blood flow, torque and the effects resulting from the unique pressure distribution on each valve.Keywords
Funding Information
- Tulsa County Heart Association
This publication has 4 references indexed in Scilit:
- Dimensions and Geometric Relationships of the Human Aortic Value as a Function of PressureCirculation Research, 1974
- An Anatomical Comparison of Human, Pig, Calf, and Sheep Aortic ValvesThe Annals of Thoracic Surgery, 1969
- Fluid Mechanics of the Aortic Root with Application to Coronary FlowNature, 1968
- Effect of Exercise on Cardiac Output, Left Coronary Flow and Myocardial Metabolism in the Unanesthetized DogCirculation Research, 1965