Detection of Methicillin-Resistant Coagulase-Negative Staphylococci by the Vitek 2 System
Open Access
- 1 September 2014
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 52 (9), 3196-3199
- https://doi.org/10.1128/jcm.01162-14
Abstract
The accurate performance of the Vitek 2 GP67 card for detecting methicillin-resistant coagulase-negative staphylococci (CoNS) is not known. We prospectively determined the ability of the Vitek 2 GP67 card to accurately detect methicillin-resistant CoNS, with mecA PCR results used as the gold standard for a 4-month period in 2012. Included in the study were 240 consecutively collected nonduplicate CoNS isolates. Cefoxitin susceptibility by disk diffusion testing was determined for all isolates. We found that the three tested systems, Vitek 2 oxacillin and cefoxitin testing and cefoxitin disk susceptibility testing, lacked specificity and, in some cases, sensitivity for detecting methicillin resistance. The Vitek 2 oxacillin and cefoxitin tests had very major error rates of 4% and 8%, respectively, and major error rates of 38% and 26%, respectively. Disk cefoxitin testing gave the best performance, with very major and major error rates of 2% and 24%, respectively. The test performances were species dependent, with the greatest errors found for Staphylococcus saprophyticus . While the 2014 CLSI guidelines recommend reporting isolates that test resistant by the oxacillin MIC or cefoxitin disk test as oxacillin resistant, following such guidelines produces erroneous results, depending on the test method and bacterial species tested. Vitek 2 cefoxitin testing is not an adequate substitute for cefoxitin disk testing. For critical-source isolates, mecA PCR, rather than Vitek 2 or cefoxitin disk testing, is required for optimal antimicrobial therapy.Keywords
This publication has 14 references indexed in Scilit:
- Evaluation of discrepancies between oxacillin and cefoxitin susceptibility in coagulase-negative staphylococciEuropean Journal of Clinical Microbiology & Infectious Diseases, 2011
- Identification of staphylococci by Phoenix: validation of a new protocol and comparison with Vitek 2Diagnostic Microbiology and Infectious Disease, 2010
- Correlation of Cefoxitin MICs with the Presence of mecA in Staphylococcus sppJournal of Clinical Microbiology, 2009
- Comparison of three phenotypic techniques for detection of methicillin resistance in Staphylococcus spp. reveals a species-dependent performanceJournal of Antimicrobial Chemotherapy, 2009
- Comparison of the MicroScan, VITEK 2, and Crystal GP with 16S rRNA sequencing and MicroSeq 500 v2.0 analysis for coagulase-negative StaphylococciBMC Microbiology, 2008
- Evaluation of the Vitek 2 System with a Variety of Staphylococcus SpeciesJournal of Clinical Microbiology, 2008
- Performances of VITEK 2 Colorimetric Cards for Identification of Gram-Positive and Gram-Negative BacteriaJournal of Clinical Microbiology, 2005
- Results of Disk Diffusion Testing with Cefoxitin Correlate with Presence of mecA in Staphylococcus sppJournal of Clinical Microbiology, 2005
- Rapid Detection of Methicillin Resistance in Coagulase-Negative Staphylococci with the VITEK 2 SystemJournal of Clinical Microbiology, 2002
- Is resistance to novobiocin a reliable test for confirmation of the identification of Staphylococcus saprophyticus?Journal of Medical Microbiology, 1989