Near Infrared Spectroscopy during pediatric cardiac surgery: errors and pitfalls

Abstract
As a result of improvements in early outcomes, long-term neurologicalal outcomes are becoming a major issue in pediatric cardiac surgery. The mechanisms of brain injury are numerous, but a vast majority of injuries are impervious to therapy and only a few are modifiable. The quality of perfusion during cardiac surgery is a modifiable factor and cerebral monitoring during bypass is the way to assess the quality of intra-operative cerebral perfusion. Near infrared spectroscopy (NIRS), as a diagnostic tool, has gained in popularity within the perfusion community. However, NIRS is becoming the standard of care before its scientific validation. This manuscript relates four clinical cases, demonstrating the limitations of NIRS monitoring during pediatric cardiac surgery as well as uncertainties about the interpretation of the recorded values. The clinical relevance of cerebral oxymetry is needed before the use of NIRS as a decision making tool. Multimodal brain monitoring with NIRS, trans-cranial Doppler and electroencephalogram are currently under way in several pediatric centers. The benefit of this time-consuming and expensive monitoring system has yet to be demonstrated.

This publication has 40 references indexed in Scilit: