Computerized image analysis for quantitative measurement of vessel diameter from cineangiograms.
- 1 August 1983
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 68 (2), 453-461
- https://doi.org/10.1161/01.cir.68.2.453
Abstract
Subjective estimates of the angiographic severity of coronary artery stenoses show variability and inaccuracy. We therefore tested the accuracy of a newly developed computerized image analysis system for quantitating vessel diameter from cineangiograms. Fourteen cylindrical phantoms of known diameter were filled with contrast medium and filmed over a wide range of clinically relevant radiographic conditions in order to develop regression equations that related computer-derived to anatomic diameters. Computer measurements of vessel diameter were unaffected by vessel size, magnification, focal spot size, thickness of scattering medium, kilovolt peak, or location within the radiographic field, but a correction factor was necessary for a small but significant (p less than .01) linear dependence on contrast medium concentration. The accuracy of computerized vessel diameter measurements ranged between +/- 59 and +/- 137 mu for all conditions except for rapid vessel motion and contrast medium concentrations of 50% or less meglumine diatrizoate (Renografin 76), both of which resulted in reduced accuracy as well as in the inability to locate lumen edges of vessels less than 1 mm in diameter.This publication has 21 references indexed in Scilit:
- A catheterization technique for reproduction of a human atherosclerotic lumen within the dog coronary artery in vivoCatheterization and Cardiovascular Diagnosis, 1983
- Analysis of the significance of scattered radiation in reduced dose mammography, including magnification effects, scatter suppression, and focal spot and detector blurringMedical Physics, 1979
- Cine‐densitometric measurement of coronary arterial stenosesCatheterization and Cardiovascular Diagnosis, 1979
- Reproducibility of a consensus panel in the interpretation of coronary angiogramsAmerican Heart Journal, 1978
- Variability in the analysis of coronary arteriograms.Circulation, 1977
- Analysis of pulsatile, viscous blood flow through diseased coronary arteries of manJournal of Biomechanics, 1977
- Field characteristics of geometric unsharpness due to the x‐ray tube focal spotMedical Physics, 1977
- Effect of asymmetric focal spots in angiographyMedical Physics, 1977
- Comparison of angiographic and postmortem findings in patients with coronary artery diseaseThe American Journal of Cardiology, 1975
- The total unsharpness in radiographyPhysics in Medicine & Biology, 1973