Computational Fluid Dynamics (CFD) Study of the 4th Generation Prototype of a Continuous Flow Ventricular Assist Device (VAD)
- 1 April 2004
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 126 (2), 180-187
- https://doi.org/10.1115/1.1688776
Abstract
The continuous flow ventricular assist device (VAD) is a miniature centrifugal pump, fully suspended by magnetic bearings, which is being developed for implantation in humans. The CF4 model is the first actual prototype of the final design product. The overall performances of blood flow in CF4 have been simulated using computational fluid dynamics (CFD) software: CFX, which is commercially available from ANSYS Inc. The flow regions modeled in CF4 include the inlet elbow, the five-blade impeller, the clearance gap below the impeller, and the exit volute. According to different needs from patients, a wide range of flow rates and revolutions per minute (RPM) have been studied. The flow rate-pressure curves are given. The streamlines in the flow field are drawn to detect stagnation points and vortices that could lead to thrombosis. The stress is calculated in the fluid field to estimate potential hemolysis. The stress is elevated to the decreased size of the blood flow paths through the smaller pump, but is still within the safe range. The thermal study on the pump, the blood and the surrounding tissue shows the temperature rise due to magnetoelectric heat sources and thermal dissipation is insignificant. CFD simulation proved valuable to demonstrate and to improve the performance of fluid flow in the design of a small size pump.Keywords
This publication has 24 references indexed in Scilit:
- Studies of Turbulence Models in a Computational Fluid Dynamics Model of a Blood PumpArtificial Organs, 2003
- Computational Fluid Dynamics Prediction of Blood Damage in a Centrifugal PumpArtificial Organs, 2003
- TRANSIENT COMPUTATIONAL FLUID DYNAMICS STUDY OF AN AXIAL BLOOD PUMPAsaio Journal, 2003
- Hematocrit, Volume Expander, Temperature, and Shear Rate Effects on Blood ViscosityAnesthesia & Analgesia, 2000
- The History of Continuous‐Flow Blood PumpsArtificial Organs, 2000
- Recommendations for Achieving Accurate Numerical Simulation of Tip Clearance Flows in Transonic Compressor RotorsJournal of Turbomachinery, 1999
- A Design Improvement Strategy for Axial Blood Pumps Using Computational Fluid DynamicsPublished by Wolters Kluwer Health ,1996
- Using Hybrid Magnetic Bearings to Completely Suspend the Impeller of a Ventricular Assist DeviceArtificial Organs, 1996
- Model for a General Mechanical Blood Damage PredictionArtificial Organs, 1995
- Deformation and fragmentation of human red blood cells in turbulent shear flowBiophysical Journal, 1975