Patient Motion in Thallium-201 Myocardial SPECT Imaging An Easily Identified Frequent Source of Artifactual Defect

Abstract
Because TI-201 SPECT requires that patients remain in an awkward position for a prolonged time, patient motion is a potentially serious source of artifactual defects on tomographic reconstructions. Thus, a simple method was developed for detection and correction of motion from SPECT images using a Co–57 point source placed on the lower anterior chest, an area remaining in the camera's field of view throughout imaging. In the absence of motion, this point source inscribes a straight line on planar summation of the 32 projections over 180°. Movement is detected by deviation from this line. The number of pixels of motion is used to shift images so that the resultant images of the point source are linear. The method of motion detection and correction was tested in 48 consecutive patients undergoing TI-201 SPECT. The corrected and uncorrected images were reconstructed and long and short axis tomographic cuts were quantitatively analyzed using circumferential profiles of maximal counts with comparison to the lower limits of normal. Motion was detected in eight of 48 patients (17%). The amount of motion was 2 pixels in three patients and 1 pixel in five patients. Quantitative defect extent was less after correction in seven of eight patients, with a mean decrease of 71% in patients with 2 pixel motion and 44% in patients with 1 pixel motion. This corresponded with a definite reduction in the size of the tomographic defect by visual analysis, and closer resemblance to quantitatively analyzed planar images performed either before or after tomography in the same patient. Patient motion is a common problem in SPECT TI-201 imaging, which may result in major tomographic artifacts. A method is described that provides a simple practical approach for the detection of patient motion. An approach to correction of motion is also described.