Clinical Experience with the Sarns Centrifugal Pump

Abstract
Since October 1986, we have had experience with 96 Sarns centrifugal pumps in 72 patients (pts). Heparinless left atrial to femoral artery or aorta bypass was used in 14 pts undergoing surgery on the thoracic aorta with 13 survivors (93%). No paraplegia or device-related complications were observed. In 57 patients, the Sarns centrifugal pump was used as a univentricular (27 pts) or biventricular (30 pts) cardiac assist device for postcardiotomy cardiogenic shock. In these patients, cardiac assist duration ranged from 2 to 434 h with a hospital survival rate of 29% in those requiring left ventricular assist and 13% in those requiring biventricular assist. Although complications were ubiquitous in this mortally ill patient population, in 5,235 pump-hours, no pump thrombosis was observed. Hospital survivors followed for 4 months to 6 years have enjoyed an improved functional class. We conclude that the Sarns centrifugal pump is an effective cardiac assist device when used to salvage patients otherwise unweanable from cardiopulmonary bypass. Partial left ventricular bypass using a centrifugal pump has become our procedure of choice for unloading the left ventricle and for maintenance of distal aortic perfusion pressure when performing surgery on the thoracic aorta. This clinical experience with the Sarns centrifugal pump appears to be similar to that reported with other centrifugal assist devices.