Fast recovery, high sensitivity NMR probe and preamplifier for low frequencies
- 1 February 1979
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 50 (2), 193-200
- https://doi.org/10.1063/1.1135786
Abstract
The design of a probe‐preamplifier combination which reduces the NMR receiver recovery time following a pulse by a factor of at least twenty at 5 MHz is described. The preamplifier, which has a noise figure of 1.3 dB, employs negative feedback to damp the probe coil (a typical effective Q is 7), but maintains the signal‐to‐noise ratio that one would expect from an undamped coil. In addition, the use of a short phase‐inverted pulse following the main transmitter pulse serves to reduce drastically the ring‐down of the probe for those high voltages where the preamplifier is inoperative. Optimum utilization of transmitter power in achieving fast pulse rise times is also discussed, as well as elimination of ’’phase‐glitch’’ by proper tuning and the use of a broadband transmitter.Keywords
This publication has 8 references indexed in Scilit:
- FET Q switch for pulsed NMRReview of Scientific Instruments, 1977
- Homogeneous rf field delay line probe for pulsed nuclear magnetic resonanceReview of Scientific Instruments, 1977
- Improved recovery time in single coil pulsed NMR spectrometersReview of Scientific Instruments, 1975
- V.H.F. f.e.t. preamplifier with 0.3 dB noise figureElectronics Letters, 1975
- A probe modification for pulsed nuclear magnetic resonance to eliminate spurious ringingJournal of Physics E: Scientific Instruments, 1974
- Coupling scheme and probe damper for pulsed nuclear magnetic resonance single coil probeReview of Scientific Instruments, 1974
- A fast recovery pulsed nuclear magnetic resonance sample probe using a delay lineReview of Scientific Instruments, 1974
- A FET analogue switch for pulsed NMR receiversJournal of Physics E: Scientific Instruments, 1973