First report on a human percutaneous transluminal implantation of a self-expanding valve prosthesis for interventional treatment of aortic valve stenosis

Abstract
Background: Percutaneous aortic valve replacement is a new technology for the treatment of patients with significant aortic valve stenosis. We present the first report on a human implantation of a self‐expanding aortic valve prosthesis, which is composed of three bovine pericardial leaflets inserted within a self‐expanding nitinol stent. The 73‐year‐old woman presented with severe symptomatic aortic valve stenosis (mean transvalvular gradient of 45 mmHg; valve area of 0.7 cm2). Surgical valve replacement had been declined for the patient because of comorbidities, including previous bypass surgery. Method and Results: A retrograde approach via the common iliac artery was used for valve deployment. The contralateral femoral vessels were used for a temporary extracorporal circulation, unloading the left ventricle during the actual stent expansion. Clinical, hemodynamic, and echocardiographic outcomes were assessed serially during the procedure. Clinical and echocardiographic follow‐up at day 1, 2, and 14 post procedure was performed to evaluate the short‐term outcome. The prosthesis was successfully deployed within the native aortic valve, with accurate and stable positioning and with no impairment of the coronary artery or vein graft blood flow. 2D and doppler echo immediately after device deployment showed a significant reduction in transaortic mean pressure gradient (from 45 to 8 mmHg) without evidence of aortic or mitral valve insufficiency. The clinical status has then significantly improved. These results remained unchanged up to the day 14 follow‐up. Conclusion: This case report demonstrates a successful percutaneous implantation of a self‐expanding aortic valve prosthesis with remarkable functional and clinical improvements in the acute and short‐term outcome.