Optical and atomic force microscopy of an explanted AcrySof intraocular lens with glistenings
- 1 April 2000
- journal article
- case report
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 26 (4), 571-575
- https://doi.org/10.1016/s0886-3350(99)00342-9
Abstract
To assess the surface morphology and cause of glistenings in an explanted AcrySof® intraocular lens (IOL). Shakai Hoken Kobe Central Hospital, Kobe, Japan. A 63-year-old Japanese man had implantation of an AcrySof IOL in the capsular bag. One month postoperatively, he had a neodymium:YAG laser capsulotomy for posterior capsule opacification, which changed the IOL’s position in the capsular bag. A few months later, the patient developed disabling night glare from intralenticular glistenings and progressive hyperopic refractive error. The IOL was explanted and then analyzed by optical microscopy and atomic force microscopy (AFM). Laboratory analysis of control AcrySof IOLs kept in a balanced salt solution at steady room and body temperature for 2 months was also performed to evaluate the cause of the glistenings observed clinically. Optical microscopy showed that the explanted AcrySof IOL had several microvacuoles; no abnormalities were observed in the control AcrySof IOLs before or after folding at the room and body temperatures. The AFM analysis showed a significant change in the surface morphology of the explanted IOL, including vacuolar formations in the posterior surface as well as numerous anterior surface irregularities. No microvacuoles or surface morphology alterations were observed in the control AcrySof IOLs by AFM analysis. The glistenings in the explanted AcrySof IOL were likely caused by temperature changes and not mechanical stress from folding.Keywords
This publication has 8 references indexed in Scilit:
- Biocompatibility of poly(methyl methacrylate), silicone, and AcrySof intraocular lenses: Randomized comparison of the cellular reaction on the anterior lens surfaceJournal of Cataract & Refractive Surgery, 1998
- Adhesiveness of AcrySof to a collagen filmJournal of Cataract & Refractive Surgery, 1998
- In vitro analysis of AcrySof intraocular lens glistenings in AcryPak and Wagon Wheel packagingJournal of Cataract & Refractive Surgery, 1998
- Human cornea and sclera studied by atomic force microscopyCell and tissue research, 1997
- Visual significance of glistenings seen in the AcrySof intraocular lensJournal of Cataract & Refractive Surgery, 1996
- Two year clinical study of a soft acrylic intraocular lensJournal of Cataract & Refractive Surgery, 1996
- Ophthalmic applications of atomic force microscopyInternational Contact Lens Clinic, 1995
- Imaging surfaces of hydrophilic contact lenses with the atomic force microscopeBiomaterials, 1993