CALIBRATION OF SOME WRIGHT PEAK FLOW METERS
- 1 April 1980
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 52 (4), 461-464
- https://doi.org/10.1093/bja/52.4.461
Abstract
The performance of Wright peak respiratory flow meters is described using an absolute and precise calibration method. This showed errors as great as 25% with the peak flow meter and 40% with the mini peak flow meter. At large flow rates errors were less. The expected variation between individual peak flow meters was about 12%. The expected variation between individual mini peak flow meters was between 16% and 22%, depending on flow rate.This publication has 1 reference indexed in Scilit: