Centrifugation as a method of improving tensile and fatigue properties of acrylic bone cement.
- 1 October 1984
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 66 (8), 1265-1273
- https://doi.org/10.2106/00004623-198466080-00016
Abstract
In this study centrifugation dramatically reduced the porosity and substantially increased the mechanical properties of bone cement. Monotonic tensile tests to failure of centrifuged specimens of cement demonstrated an increase of 24 per cent in the mean ultimate tensile strength compared with the control value. Mean ultimate tensile strain was improved by 54 per cent. In fully reversed tension-compression fatigue-testing, centrifugation resulted in a mean increase in fatigue life of 136 per cent. These strong advantages in mechanical properties were obtained without any detrimental changes. There was no change in elastic modulus, setting time, or peak temperature. Handling properties were improved. There was no increase in systemic toxicity as demonstrated in dogs by assessment of arterial blood-pressure response and peak levels of monomer in the serum during simulated total hip arthroplasty. We also present a practical system of cement centrifugation and delivery that is suitable for use in the operating room.This publication has 13 references indexed in Scilit:
- A ten-year follow-up of one hundred consecutive Müller curved-stem total hip-replacement arthroplasties.Journal of Bone and Joint Surgery, 1982
- Ten-year follow-up study of total hip replacement.Journal of Bone and Joint Surgery, 1982
- Total hip replacement in patients younger than thirty years old. A five-year follow-up study.Journal of Bone and Joint Surgery, 1981
- A model femur for in vitro testing of femoral componentsJournal of Biomedical Engineering, 1981
- A ten-year follow-up study of our first one hundred consecutive Charnley total hip replacements.Journal of Bone and Joint Surgery, 1981
- An analysis of femoral component stem design in total hip arthroplasty.Journal of Bone and Joint Surgery, 1980
- Improved fixation of the femoral component after total hip replacement using a methacrylate intramedullary plug.Journal of Bone and Joint Surgery, 1978
- Fracture characteristics of acrylic bone cements. II. FatigueJournal of Biomedical Materials Research, 1977
- On the fate of methyl methacrylate in bloodJournal of Biomedical Materials Research, 1977
- Carbon fiber‐reinforced bone cement in orthopedic surgeryJournal of Biomedical Materials Research, 1976