Method and a simple apparatus for rapid simultaneous measurement of resonance frequency and Q factor of a quartz crystal
- 1 May 1996
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 67 (5), 1961-1964
- https://doi.org/10.1063/1.1146951
Abstract
The quartz crystal microbalance is finding many novel uses beyond its traditional utility as a thin film deposition controller. In many situations where the quartz crystal is immersed in a viscous fluid, it is desirable to measure the crystal’s Q factor in addition to its resonance frequency. Here, we describe a simple method for simultaneous measurement of the resonance frequency, the quality factor, and the activity or relative amplitude of oscillation of a quartz crystal in vacuum and in gaseous environment. The hardware, comprising of a synthesized sine wave generator and a phase detector, passively interrogates the crystal to determine its resonance frequency and activity. The passive measurement system can detect a resonance even when the crystal is heavily damped. Representative measurement results for an AT‐cut quartz crystal in a number of gases at various pressures are presented. The methodology described here may be readily applied to quartz crystals immersed in a liquid.Keywords
This publication has 11 references indexed in Scilit:
- Application of the quartz crystal microbalance to measurement of the concentration of electrolyte in lead/acid batteriesJournal of Power Sources, 1995
- Quartz crystal microbalance setup for frequency and Q-factor measurements in gaseous and liquid environmentsReview of Scientific Instruments, 1995
- Thermophysical Properties of Penetrants in Polymers via a Piezoelectric Quartz Crystal MicrobalanceJournal of Chemical & Engineering Data, 1995
- Application of crown ether coated piezoelectric crystal as a detector for ion chromatographyThe Analyst, 1995
- Recent developments in piezoelectric immunosensors. A reviewThe Analyst, 1994
- Improved methodology for testing and characterization of piezodelectric gas sensorsAnalytica Chimica Acta, 1992
- Improving the accuracy of a quartz crystal microbalance with automatic determination of acoustic impedance ratioReview of Scientific Instruments, 1991
- Nanotribology of a Kr monolayer: A quartz-crystal microbalance study of atomic-scale frictionPhysical Review Letters, 1991
- Quartz crystal microbalance thin-film dissolution rate monitorReview of Scientific Instruments, 1989
- An analysis of contoured crystal resonators operating in overtones of coupled thickness shear and thickness twistThe Journal of the Acoustical Society of America, 1979