Value of Automatic Processing and Reliability of Stored Data in an Implanted Pacemaker: Initial Results in 59 Patients
- 1 December 1997
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 20 (12), 2893-2898
- https://doi.org/10.1111/j.1540-8159.1997.tb05457.x
Abstract
Stored data in implantable pacemakers have rarely been used as a diagnostic tool because of the complexity. Our group has developed software called AIDA, providing an automatic interpretation of data stored in memories of the Chorus (ELA medical) pacemaker. We compared the results of AIDA analysis to surface ECG Holter interpretation in 59 patients (age 75 +/- 9 years). In 33 cases, neither AIDA nor the Holter found any anomaly. Eleven patients demonstrated episodes of supraventricular tachycardia (SVT), confirmed by AIDA in ten patients; AIDA failure was due to nonsustained episodes of SVT not inducing mode switch. Loss of atrial sensing, pacemaker-mediated tachycardia, and ventricular extrasystoles were detected by AIDA in ten patients. Traditional Holter missed three cases. This initial study confirms that stored pacemaker data, automatically interpreted can provide reliable information over a 24-hour period.Keywords
This publication has 5 references indexed in Scilit:
- Natural history of atrial arrhythmia in DDD paced patients using the permanent holter function of the deviceJournal of the American College of Cardiology, 1996
- Diagnosis of Atrial Arrhythmias Using the Holter Function of a New DDD PacemakerPacing and Clinical Electrophysiology, 1994
- A New Algorithm to Solve Endless Loop Tachycardia in DDD Pacing: A Multi‐Center Study of 91 PatientsPacing and Clinical Electrophysiology, 1990
- Utility of Rate Histograms in Programming and Follow-Up of a DDDR PacemakerMayo Clinic Proceedings, 1989
- Data Storage and Retrieval by Implantable Pacemakers for Diagnostic PurposesPacing and Clinical Electrophysiology, 1984