A Realistic Computer-Simulated Brain Phantom for Evaluation of PET Charactenstics

Abstract
To evaluate accurately the imaging characteristics of positron emission tomography (PET), a realistic computer-simulated brain phantom was developed. A cross-sectional slice from a human cadaver brain was chosen for its combination of gray matter, white matter, and cerebrospinal fluid (CSF) regions. The slice was photographed and digitized into a gray-level image with a video digitizer, boundary edges were located around cerebral structures in the digitized image, and each structural region was assigned a uniform pixel value dependent on both the cerebral parameter (e.g., blood flow, oxygen uptake, metabolic rate) under investigation and the type of structure (gray matter, white matter, CSF). Line integrals through the regions were generated at various angular and transverse positions according to specific physical characteristics (such as detector line-spread function) of a tomographic scanner configuration to create a set of simulated but realistic projection measurements. The set of projection measurements can be processed with any standard reconstruction program to create a tomographic image to reveal the effects of various PET characteristics. Investigations with this computer-simulated brain phantom have demonstrated its usefulness for examining the interrelations among neuroanatomical structure volume, tomographic spatial resolution, partial volume effect, and nonlinear parameter estimation. Transportability of the simulated phantom and the procedure to other medical imaging environments is described, and limitations of this simulation procedure are discussed.