Evaluation of the pathotec Rapid I-D system for identification of Enterobacteriaceae
- 1 April 1975
- journal article
- clinical trial
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 1 (4), 359-362
- https://doi.org/10.1128/jcm.1.4.359-362.1975
Abstract
The PathoTec Rapid I-D System for identifying Enterobacteriaceae was evaluated with 471 cultures. In 4,910 individual test comparisons, 95.5% of the results agreed, with results of only two test strips, those for esculin hydrolysis and urease production, agreeing with conventional tests in less than 94% of the trials. The PathoTec system exhibited 94.3% accuracy in identifying these cultures in a double-blind study with conventional media and procedures as the alternate system. Two newly developed test strips, for 0-nitrophenyl-beta-D-galactopyranoside and ornithine decarboxylase, were found to be highly reliable.Keywords
This publication has 4 references indexed in Scilit:
- Evaluation of the PathoTec "Rapid I-D System".1973
- Evaluation of the PathoTec "Rapid I-D System" and two additional Experimental reagent-impregnated paper strips.1973
- Evaluation of reagent-impregnated strips for identification of Enterobacteriaceae.1971
- Comparative study of the efficacy of seven paper-reagent strips and conventional biochemical tests in identifying gram-negative organisms.1969