Evaluation of the pathotec Rapid I-D system for identification of Enterobacteriaceae

Abstract
The PathoTec Rapid I-D System for identifying Enterobacteriaceae was evaluated with 471 cultures. In 4,910 individual test comparisons, 95.5% of the results agreed, with results of only two test strips, those for esculin hydrolysis and urease production, agreeing with conventional tests in less than 94% of the trials. The PathoTec system exhibited 94.3% accuracy in identifying these cultures in a double-blind study with conventional media and procedures as the alternate system. Two newly developed test strips, for 0-nitrophenyl-beta-D-galactopyranoside and ornithine decarboxylase, were found to be highly reliable.