Integrated circuit ac mutual inductance bridge for magnetic susceptibility measurements
- 1 September 1978
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 49 (9), 1279-1281
- https://doi.org/10.1063/1.1135567
Abstract
A versatile ac mutual inductance bridge is described which uses integrated circuit operational amplifiers (op-amps) for the measurement of magnetic susceptibilities of ferro-, ferri-, and para-magnetic samples. The circuit employs op-amps both for balancing the bridge and for detecting the differential signal from the sample coil. The response of the circuit is linear over a wide range of sample susceptibilities, and is calibrated directly in absolute units (emu) using a multiturn digital potentiometer. The sensitivity of the instrument for weak paramagnetic samples (1 g) is +/-3x10(-7) emu.Keywords
This publication has 5 references indexed in Scilit:
- Mutual Inductance Bridge for ac Susceptibility Measurements at Low FrequenciesReview of Scientific Instruments, 1965
- Magnetic Susceptibility: Trends in Instrumentation, Research, and ApplicationsAnalytical Chemistry, 1964
- Electronic ac Mutual Inductance Bridge for Measuring Small Susceptibilities at Low TemperaturesReview of Scientific Instruments, 1958
- Installation for adiabatic demagnetization experiments at the National Bureau of StandardsJournal of Research of the National Bureau of Standards, 1954
- A precision method for the comparison of unequal mutual inductances at telephonic frequenciesJournal of Scientific Instruments, 1925