A Chemical-Stop Rapid Flow Apparatus
- 1 March 1971
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 42 (3), 356-358
- https://doi.org/10.1063/1.1685093
Abstract
A rapid flow apparatus with multiple mixers is described, capable of measuring reaction times of 0.008 to 0.5 sec. The solution‐containing parts are of Teflon and Lucite to minimize metal contamination. A rack and pinion drive, using an electric clutch, is described, along with a positive assay valve needed when the apparatus is used as a chemical‐stop flow apparatus.Keywords
This publication has 4 references indexed in Scilit:
- Transient state phosphate production in the hydrolysis of nucleoside triphosphates by myosinBiochemistry, 1970
- Apparatus for rapid and sensitive spectrophotometryBiochemical Journal, 1964
- Mixing Machine of High Precision for Study of Rapid ReactionsReview of Scientific Instruments, 1955