Contrast enhancement of high-energy radiotherapy films
- 1 November 1979
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 52 (623), 880-887
- https://doi.org/10.1259/0007-1285-52-623-880
Abstract
An order-of-magnitude improvement in the contrast of high-energy localization and verification films has been achieved through the application of a simple, inexpensive, contrast enhancement technique. The method involves making reversal contact prints of the original film onto ordinary X-ray film with equipment commonly available in any radiotherapy department. This results in .gamma. multiplication. The theory as well as several applications of this effect are presented.This publication has 2 references indexed in Scilit:
- Improved Technique for Obtaining Teletherapy Portal Radiographs with High-Energy PhotonsRadiology, 1978