Perturbation correction factors for some ionization chambers commonly applied in electron beams
- 1 December 1991
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 36 (12), 1639-1652
- https://doi.org/10.1088/0031-9155/36/12/008
Abstract
Values for the fluence perturbation correction factor were determined in a number of electron beams for the commonly applied PTW/Markus plane-parallel chamber and the cylindrical NE 2571 Farmer chamber. These data were determined relative to the NACP plane-parallel chamber as a function of the mean energy at depth, Ez, calculated according to the method recommended in modern dosimetry protocols. For the cylindrical chamber the results are in agreement with the data recommended in these protocols. For the PTW/Markus chamber a small but significant fluence perturbation correction is necessary: up to about 3% for a value of Ez=2 MeV. In order to check the value of these fluence perturbation correction values, dose measurements were performed ionometrically and compared with values determined with the Fricke dosimetry system.Keywords
This publication has 17 references indexed in Scilit:
- Comparison of ferrous sulfate (Fricke) and ionization dosimetry for high‐energy photon and electron beamsMedical Physics, 1990
- Measurement of replacement factors for a parallel-plate chamberMedical Physics, 1989
- Replacement correction factors for electron measurements with a parallel-plate chamberMedical Physics, 1987
- Determinations of N gas and P repl factors from commercially available parallel-plate chambers: AAPM Task Group 21 protocolMedical Physics, 1986
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983
- Ferrous Sulphate Dosimeter for Control of Ionization Chamber Dosimetry of Electron and60Co Gamma BeamsActa Radiologica: Oncology, 1982
- Mean energy in electron beamsMedical Physics, 1981
- Calibration and use of Plane-Parallel Ionization Chambers for the Determination of Absorbed Dose in Electron BeamsActa Radiologica: Oncology, 1981
- Current collection and ionic recombination in small cylindrical ionization chambers exposed to pulsed radiationThe British Journal of Radiology, 1980
- Acidic potassium dichromate solutions as ultraviolet absorbance standardsJournal of Research of the National Bureau of Standards Section A: Physics and Chemistry, 1976