Streak contrast in field-ion micrographs
- 1 May 1966
- journal article
- other
- Published by Taylor & Francis in Philosophical Magazine
- Vol. 13 (125), 1085-1087
- https://doi.org/10.1080/14786436608213159
Abstract
Asymmetric field-ion microscope specimens have stable end-forms which are themselves asymmetric and cannot be field-evaporated to give an undistorted field-ion image. On such specimens surface steps may persist during field evaporation and give rise to image streaking through field-focusing of the ion beams from the step region. Persistent steps on asymmetric tips arise from prior cold work or deformation during the tip preparation process. Faulty electropolishing, which is responsible for the tip asymmetry, is thus the primary cause of streak contrast.Keywords
This publication has 1 reference indexed in Scilit: