Laboratory evaluation of the Anderson Technical semiautomated susceptibility testing system
Open Access
- 1 December 1981
- journal article
- research article
- Published by American Society for Microbiology in Antimicrobial Agents and Chemotherapy
- Vol. 20 (6), 777-782
- https://doi.org/10.1128/aac.20.6.777
Abstract
Anderson Technical, Inc. has developed rapid, semiautomated equipment for antimicrobial susceptibility testing. A total of 310 fresh clinical bacterial isolates (226 gram-negative and 75 gram-positive) were tested with the Anderson Technical system and compared with those of Micro-Media Systems, Inc. For the gram-positive organisms, 74.3% of the test pairs had identical minimum inhibitory concentration values, whereas 99.1 and 0.9% of the test pairs had minimal inhibitory concentration values differing by less than or equal to 1 and greater than 1 dilution level, respectively. Identical minimum inhibitory concentration values were obtained for 67.2% of the gram-negative test pairs, whereas 97.6 and 2.4% differed by less than or equal to 1 and greater than 1 dilution level, respectively. For all organisms tested, 98.0% differed by less than or equal to 1 dilution level. The Anderson Technical equipment proved to be a rapid and flexible system for microdilution testing.This publication has 7 references indexed in Scilit:
- Evaluation of the Micro-Media System for Quantitative Antimicrobial Drug Susceptibility Testing: a Collaborative StudyAntimicrobial Agents and Chemotherapy, 1978
- Laboratory Evaluation of a Rapid, Automated Susceptibility Testing System: Report of a Collaborative StudyAntimicrobial Agents and Chemotherapy, 1975
- A Microdilution Method for Antibiotic Susceptibility Testing: An EvaluationAmerican Journal of Clinical Pathology, 1970