Measurement of absolute tracer concentrations in tissue sections by using digital imaging fluorescence microscopy. Application to the study of plasma protein uptake by the arterial wall
- 1 February 1994
- journal article
- Published by Wiley in Journal of Microscopy
- Vol. 173 (2), 127-141
- https://doi.org/10.1111/j.1365-2818.1994.tb03435.x
Abstract
SUMMARY: Digital imaging fluorescence microscopy (DIFM) of tissue sections was used to quantify uptake of labelled plasma proteins by the arterial wall. Several aspects of the measuring system were investigated so that absolute tracer concentrations and their local variation could be derived from digitized images. These investigations may be relevant to other studies employing DIFM.Nonlinearities were found to arise from offsets in the video digitizers, from background fluorescence and stray light within the microscope and from the transfer characteristics of the intensified CCD camera. Camera gain controls showed complex behaviour. Camera output fell substantially for several hours after switching on and was affected by room temperature. Large spatial variations in response were caused by the geometry of the microscope optics and by the image intensifier. However, the ratios between areas were not affected by light intensity or camera gain settings. Measured intensities were independent of the depthwise location of fluorophores within tissue sections but they were affected by the emission from objects outside the measuring area. Photobleaching of tracer varied significantly over the range of excitation intensities and durations used but was not concentration dependent. Methods used to correct these effects and obtain a uniform, linear and constant relationship between concentration and grey level are described.Using the system and appropriate corrections, in vivo uptake of sulphorhodamine‐B‐labelled serum albumin by the rabbit aortic wall was investigated. Results obtained for the mean uptake of tracer and its local variation were quantitatively similar to those previously obtained with nonmicroscopic methods.Keywords
This publication has 16 references indexed in Scilit:
- Chapter 2 Validation of an Imaging System: Steps to Evaluate and Validate a Microscope Imaging System for Quantitative StudiesMethods in Cell Biology, 1989
- Multiple spectral parameter imaging in quantitative fluorescence microscopy. I: Quantitativn of bead standardsComputerized Medical Imaging and Graphics, 1989
- Quantification in vivo of increased LDL content and rate of LDL degradation in normal rabbit aorta occurring at sites susceptible to early atherosclerotic lesions.Circulation Research, 1988
- Illumination and detection systems for quantitative fluorescence microscopyJournal of Microscopy, 1987
- Video MicroscopyPublished by Springer Nature ,1986
- Net albumin transport across the wall of the rabbit common carotid artery perfused in situAtherosclerosis, 1980
- The distribution of labeled albumin across the rabbit thoracic aorta in vivo.Circulation Research, 1977
- Line Photometry and the Influence of the Eberhard EffectJournal of the Optical Society of America, 1968
- Antigenic analysis of Entamoeba histolytica by means of fluorescent antibody II. E. histolytica and E. hartmanniExperimental Parasitology, 1960
- FLUORESCENT PROTEIN TRACERS A SIMPLE ALTERNATIVE TO FLUORESCEINThe Lancet, 1958