The point of epiphyseal penetration affects rotational stability of screw fixation in slipped capital femoral epiphysis: A biomechanical study
- 19 May 2020
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 38 (12), 2634-2639
- https://doi.org/10.1002/jor.24747
Abstract
The epiphyseal tubercle, a posterosuperior projection of the epiphysis into the metaphysis, serves as the axis of rotation in slipped capital femoral epiphysis (SCFE) and a source of physeal stability. We hypothesized that in a biomechanical model of single screw fixation of stable SCFE, a screw passing through the epiphyseal tubercle (the axis of rotation) would confer less rotational stability than a centrally placed screw. Three femurs were selected from a sample population of 8‐ to 15‐year‐old healthy hips to represent three stages of maturation: a “young” femur with a prominent epiphyseal tubercle and decreased epiphyseal cupping around the metaphysis, a “median” femur with a subsiding tubercle, and a “mature” femur with a subsided epiphyseal tubercle and increased peripheral epiphyseal cupping. Specimens were three‐dimensional printed with one of two screw trajectories: passing centrally in the epiphysis or directly through the epiphyseal tubercle. Resistance to rotational displacement was measured through stiffness and maximum torque over 30° degrees of displacement. In the “young” model, epiphyseal tubercle screw position conferred less rotational stiffness and required less maximum torque during rotational displacement when compared to a centrally placed screw (P < .001). In the “median” and “mature” models where the tubercle has subsided and is replaced by peripheral epiphyseal cupping, screw position through the tubercle was associated with equal or greater rotational stiffness and maximum torque during displacement as a centrally placed screw.Keywords
This publication has 22 references indexed in Scilit:
- Biomechanical Comparison of Perpendicular Versus Oblique In Situ Screw Fixation of Slipped Capital Femoral EpiphysisJournal of Pediatric Orthopaedics, 2015
- Lateral insertion is a good prognostic factor after in situ fixation in slipped capital femoral epiphysisBMC Musculoskeletal Disorders, 2014
- Mechanical failure of single cannulated screw fixation of unstable slipped capital femoral epiphysisJournal of Pediatric Orthopaedics B, 2014
- An Anatomic Study of the Epiphyseal Tubercle and Its Importance in the Pathogenesis of Slipped Capital Femoral EpiphysisJournal of Bone and Joint Surgery, 2013
- Biomechanical Study of 16-mm Threaded, 32-mm Threaded, and Fully Threaded SCFE Screw FixationJournal of Pediatric Orthopaedics, 2012
- The epiphyseal tubercle in adolescent hipsActa Orthopaedica, 2009
- Biomechanical Comparison of Fully and Partially Threaded Screws for Fixation of Slipped Capital Femoral EpiphysisJournal of Pediatric Orthopaedics, 2008
- Does the upper femoral epiphysis slip or rotate?The Journal of Bone and Joint Surgery. British volume, 2007
- Biomechanical Analysis of Single Screw Fixation for Slipped Capital Femoral EpiphysisJournal of Pediatric Orthopaedics, 2006