A simulation of the effects of set-up error and changes in breast volume on conventional and intensity-modulated treatments in breast radiotherapy
- 17 April 2001
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 46 (5), 1451-1471
- https://doi.org/10.1088/0031-9155/46/5/309
Abstract
The effect of interfractional patient movement on dosimetry has been investigated for breast radiotherapy. Errors in patient set-up and changes in breast volume were simulated individually to determine how each contributes to the total dosimetric error. Two treatment techniques were investigated: a conventional treatment and an intensity-modulated treatment delivered using compensators. Six patients were investigated and anterior-posterior (AP) and superior-inferior (SI) displacements were simulated by displacing the isocentre in both directions by 2, 5 and 10 mm. A model of the breast was developed from the six patients to simulate changes in breast volume. In this model, the breast was described as a set of semi-ellipses. The volume of the breast was changed by varying the magnitude of the semi-major and semi-minor axes. Anisotropic changes in breast volume were also investigated. The dosimetric error was evaluated for each dose plan by calculating the volume outside the 95-105% dose range resulting from the simulations. A number of parameters describing the size and shape of the breast were also investigated to determine whether a susceptibility of outline sets to interfractional patient movement could be predicted. A parameter describing the increase in the breast volume outside the 95-105% dose range was calculated for AP aKeywords
This publication has 24 references indexed in Scilit:
- The dosimetric consequences of inter-fractional patient movement on conventional and intensity-modulated breast radiotherapy treatmentsRadiotherapy and Oncology, 2000
- Intensity-modulated tangential beam irradiation of the intact breastInternational Journal of Radiation Oncology*Biology*Physics, 1999
- The use of compensators to optimise the three dimensional dose distribution in radiotherapy of the intact breastRadiotherapy and Oncology, 1999
- Practical implementation of compensators in breast radiotherapyRadiotherapy and Oncology, 1998
- Impact of setup variability on incidental lung irradiation during tangential breast treatmentInternational Journal of Radiation Oncology*Biology*Physics, 1997
- Intra- and interfractional reproducibility of tangential breast fields: A prospective on-line portal imaging studyInternational Journal of Radiation Oncology*Biology*Physics, 1996
- Design of compensators for breast radiotherapy using electronic portal imagingRadiotherapy and Oncology, 1995
- Treatment delivery accuracy in intensity-modulated conformal radiotherapyPhysics in Medicine & Biology, 1995
- A study of effective attenuation coefficient for calculating tissue compensator thicknessMedical Physics, 1990
- Computer aided design and verification of megavoltage tissue compensators for oblique beamsMedical Physics, 1988