A two‐step scheme for distortion rectification of magnetic resonance images
- 12 August 2009
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 36 (9Part1), 3917-3926
- https://doi.org/10.1118/1.3180107
Abstract
The aim of this work is to demonstrate a complete, robust, and time-efficient method for distortion correction of magnetic resonance (MR) images. It is well known that MR images suffer from both machine-related spatial distortions [gradient nonlinearity and main field (B-0) inhomogeneity] and patient-related spatial distortions (susceptibility and chemical shift artifacts), and growing interest in the area of MR-based radiotherapy treatment planning has put new requirements on the geometric accuracy of such images. The authors present a two-step method that combines a phantom-based reverse gradient technique for measurement of gradient nonlinearities and a patient-based phase difference mapping technique for measurement of B-0 inhomogeneities, susceptibility, and chemical shift distortions. The phase difference mapping technique adds only minutes to the total patient scan time and can be used to correct a variety of images of the same patient and anatomy. The technique was tested on several different phantoms, each designed to isolate one type of distortion. The mean distortion was reduced to 0.2 +/- 0.1 mm in both gradient echo and spin echo images of a grid phantom. For the more difficult case of a highly distorted echo planar image, residual distortion was reduced to subvoxel dimensions. As a final step, the technique was implemented on patient images. The current technique is effective, time efficient, and robust and provides promise for preparing distortion-rectified MR images for use in MR-based treatment planning. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3180107]Keywords
This publication has 26 references indexed in Scilit:
- Phase labeling using sensitivity encoding (PLUS): Data acquisition and image reconstruction for geometric distortion correction in EPIMagnetic Resonance in Medicine, 2008
- Concurrent correction of geometric distortion and motion using the map-slice-to-volume method in echo-planar imagingMagnetic Resonance Imaging, 2008
- A study on the magnetic resonance imaging (MRI)-based radiation treatment planning of intracranial lesionsPhysics in Medicine & Biology, 2008
- Distortion-correctedT2weighted MRI: a novel approach to prostate radiotherapy planningThe British Journal of Radiology, 2007
- Magnetic Resonance–Based Treatment Planning for Prostate Intensity-Modulated Radiotherapy: Creation of Digitally Reconstructed RadiographsInternational Journal of Radiation Oncology*Biology*Physics, 2007
- Feasibility of using MRI alone for 3D Radiation Treatment Planning in Brain TumorsJapanese Journal of Clinical Oncology, 2007
- Correction for geometric distortion and N/2 ghosting in EPI by phase labeling for additional coordinate encoding (PLACE)Magnetic Resonance in Medicine, 2007
- A complete distortion correction for MR images: II. Rectification of static-field inhomogeneities by similarity-based profile mappingPhysics in Medicine & Biology, 2005
- A complete distortion correction for MR images: I. Gradient warp correctionPhysics in Medicine & Biology, 2005
- A technique for accurate magnetic resonance imaging in the presence of field inhomogeneitiesIEEE Transactions on Medical Imaging, 1992