Adaptive temporal resolution optimization in helical cardiac cone beam CT reconstruction
- 18 November 2003
- journal article
- clinical trial
- Published by Wiley in Medical Physics
- Vol. 30 (12), 3072-3080
- https://doi.org/10.1118/1.1624756
Abstract
Cone beam computed tomography scanners in combination with heart rate adaptive reconstruction schemes have the potential to enable cardiac volumetric computed tomography (CT) imaging for a larger number of patients and applications. In this publication, an adaptive scheme for the automatic and patient‐specific reconstruction optimization is introduced to improve the temporal resolution and image quality. The optimization method permits the automatic determination of the required amount of gated helical cone beam projection data for the reconstruction volume. It furthermore allows one to optimize subvolume reconstruction yielding an increased temporal resolution. In addition, methods for the assessment of the temporal resolution are given which enable a quantitative documentation of the reconstruction improvements. Results are presented for patient data sets acquired in low pitch helical mode using a 16‐slice cone beam CT system with parallel ECG recording.Keywords
This publication has 18 references indexed in Scilit:
- Theoretically Exact Filtered Backprojection-Type Inversion Algorithm for Spiral CTSIAM Journal on Applied Mathematics, 2002
- Artifact analysis of approximate helical cone‐beam CT reconstruction algorithmsMedical Physics, 2001
- ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heartMedical Physics, 2000
- High temporal resolution for multislice helical computed tomographyMedical Physics, 2000
- Multislice helical CT: image temporal resolutionIEEE Transactions on Medical Imaging, 2000
- Exact (spiral+circles) scan region-of-interest cone beam reconstruction via backprojectionIEEE Transactions on Medical Imaging, 2000
- ECG-correlated imaging of the heart with subsecond multislice spiral CTIEEE Transactions on Medical Imaging, 2000
- Quasi-exact filtered backprojection algorithm for long-object problem in helical cone-beam tomographyIEEE Transactions on Medical Imaging, 2000
- Electrocardiogram‐correlated image reconstruction from subsecond spiral computed tomography scans of the heartMedical Physics, 1998
- An Inversion Formula for Cone-Beam ReconstructionSIAM Journal on Applied Mathematics, 1983